DescriptionThe All-rounder amongst Polarimeter with continuous measurement, resolution ±0,001°
Fully automatic circle polarimeter Polartronic M-Series
Polarimeter der M-Serie are high resolution polarimeters with following features: - Accurate measuring of the optical activity of solutions according to ICUMSA, O.I.M.L., Australian Standard K157, European and American Pharmacopeia
- Direct reading of the rotation angle without mechanical transmission is resulting in guaranteed unmatchable precision over the entire measuring range
- Recommended for highest requirements in quality and research laboratories
- Short measuring times, no mechanical abrasion
- The measurement is carried out constantly without pressing a button so that flow through measurements can easily be performed
- The customer can program up to 10 special scales (glucose, mono sodium glutamate or other) and 10 methods
- Automatic temperature compensation if using appropriate polarimeter tubes
- Indication of the optical density
- Serial to Ethernet converter for data transfer to PC
- The instrument has a versatile software for test routines and configurations
- GLP/GMP conform documentation
- 21 CFR part 11 ready, appropriate software "Aquisys 2008" optionally available
- High quality "Made in Germany", consultation and service at site possible
In conjunction with the Schmidt+Haensch Software AQUISYS 2008 all requirements of 21 CFR Part 11 will be performed.
AQUISYS 2008 software is 21 CFR Part 11 compliant for Schmid+Haensch's instruments.

- Audit Trail function (history logging)
- Report generation
- Electronic signature
- Encrypted electronic data storage and archiving
- Data transfer to Excel
- Remote modus
Technical detailsPolartronic M 100 (replacement for Polartronic MHZ-8) | | Measuring range: | ±360° / ± 259°Z | Resolution: | 0,001° / 0,01°Z | Precision: | ± 0,005° / ± 0,015°Z over the entire measuring range
± 0.002° between ± 1° | Light source: | LED, life time guarantee | Wavelength: | 589 nm | Sensitivity
Polarimeter tube: | up to OD 5
up to 200 mm length | Temperature correction: | automatic with appropriate tubes | Response time: | approx. 4 sec | Remote-operation: | possible | Display: | Graphics-LCD 16 x 16 characters | Interfaces: | 2 x RS 232 C, 1 x parallel, 1 x PS2, 1 x USB /Ethernet(optional) | Power supply: | 110 / 220 V / 50 / 60 Hz | | | Weight: | ca. 28 kg | Standards: | ICUMSA, O.I.M.L., Australian Standard K 157 |
Polartronic M 101 |
difference | Wavelength: | 882nm | Polartronic M 202 | difference | Wavelength: | 589nm and 882nm |
|
| | |
|
| | |
|